Alpha pH 2000D pH/ORP controller/transmitter

Seite 2 - Preface

2.3 Unit Overview Wall mounting version - 5 -

Seite 3 - Safety Information

2.3.1 Display Overview The LC display shows two alpha-numerical fields for parameters and measured values as well as various mode and status ind

Seite 4 - TABLE OF CONTENTS

2.3.3 LED indicators Relay indicators If REL key is pressed the LED (A, B or W) indicates to which Relay (A, B or Wash) the displayed limit

Seite 5

2.3.5 Menu overview CCD “000” = Check calibration parameters (View only mode) CCD “11” = Calibration mode SCD “000” = Check setup parameters

Seite 6 - - 1

3. Assembly and Installation 3.1 Mounting the Unit Wall mounting version 111.50 [4.39]144 [5.67]Un

Seite 7 - 1.3 Safety Instructions

3.2 Connection Diagram Caution: Ensure electrical mains are disconnected before proceeding. Connections for wall mounting version 1L2NCurr

Seite 8 - - 3

4 Normal Operation 4.1 Measurement Mode When the controller/transmitter is powered on, the display first shows all segments briefly, after which

Seite 9

5 Calibration Mode You can access the Calibration mode directly from the Measurement mode by pressing the CAL key and entering the Calibration se

Seite 10 - 2.3 Unit Overview

5.2 pH Calibration If the unit of measure is in pH (default), then the calibration is also carried out in pH. This transmitter features a one-poi

Seite 11 - 2.3.1 Display Overview

5.3 ORP-mV Calibration If controller is configured for ORP-mV measuring you can perform a one-point calibration. MEASCALCALHOLDCALHOLDmVmVR.mVCAL

Seite 12 - - 7

Preface This manual serves to explain the use of the Alpha-pH2000D series pH/ORP controller/transmitter. The manual functions in two ways, firstly a

Seite 13

5.5 View actual zero and slope MEASPHOCATCCALENTpHCALENT 1. While in Measurement mode press the CAL key. The display will prompt you to enter a

Seite 14 - - 9

6 Setup Mode 6.1 Enter Setup mode In the Setup mode the transmitter can be configured to your individual requirements. 1. While in Measurement

Seite 15 - Time Delay

6.2 Electrode Offset (OFS) sub-function The electrode offset sub-function is only available if the transmitter is configured for pH measuring. Us

Seite 16 - 4 Normal Operation

6.3 Setting temperature (Set °C°F) sub-function SETUPHOLDSETUPHOLDOCSETUPHOLDSETUPHOLDSETUPHOLDOOCFSETUPHOLDOCATCSETUPHOLDSETUPHOLDENTENTENTENT

Seite 17 - EN TEREN TER

6.4 Control Relay A / Relay B (SP1/SP2) sub-function The SP1 sub-function determines the operating parameters for Relay A; while SP2 defines th

Seite 18

6. Setting the off-delay time lag: press the ▲ or ▼ key to enter the off-delay time for set point 1 (set point 2). Your controller will delay de

Seite 19

6.5 Controller (CNTR) sub-function The CNTR sub-function determineshe controller’s parameters. SETUP SETUP HOLD HOLD ENT SETUP HOLD ENT ENT E

Seite 20 - - 15

2. Selecting the controller type: press the ▲ or ▼ key to select the suitable controller type: − OFF = controller off Use control Off to operat

Seite 21 - 6 Setup Mode

6.6 Current Output 1 sub-function In this sub-function you set the current output range of the transmitter for pH or ORP values. 22334455SETU

Seite 22 - - 17

6.7 Current Output 2 sub-function In this sub-function you set the current output range of the transmitter for temperature values. ENTENTENTE

Seite 23 - - 18

Safety Information The Controller/ Transmitter shall be installed and operated only in the manner specified in the Instruction manual. Only skilled,

Seite 24

6.8 Wash relay (WASH) sub-function In this sub-function you set the parameters for the wash relay. 1223345ENTENT ENTENTENTENTENTSETUPHOLDSETU

Seite 25 - - 20

6.9 Configuration (CNFG) sub-function In this sub-function you configure the transmitter to suit your requirements. SETUP HOLD SETUP HOLD SET

Seite 26

Note: With this parameter you select whether the alarm contact will operate as a steady contact or a fleeting (single pulse) contact. Pulse conta

Seite 27 - - 22

7 Relay Modes You can control devices connected to Relay A, Relay B or wash relay via the front panel of the transmitter. In Automatic mode, the

Seite 28

7.2 Manual relay mode In manual relay mode, you can manually turn “on” and “off” the control devices connected to Relay A, Relay B or Wash relay

Seite 29 - - 24

8 Technical Specifications 8.1 General Specification (a) pH pH Measuring Range -2.00 to 16.00 pH Resolution 0.01 pH Accuracy ± 0.01 pH (b) m

Seite 30

8.4 Alarm Function – Alarm Relay Alarm Delay 0 to 2000 seconds Alarm control Steady or Fleet (pulse) Contact Outputs One SPDT relay Switchi

Seite 31

8.11 Electromagnetic Compliance (EMC) Specifications Emitted Interference EN 61326 Immunity to Interference EN 61326 8.12 Environmental Cond

Seite 32 - - 27

9 Accessories 9.1 Replacement Unit Product Description Eutech Instruments Order Code Alpha pH 2000 Controller / Transmitter, wall mount version

Seite 33 - 7 Relay Modes

10 General Information 10.1 Warranty Eutech Instruments/Oakton Instruments warrant this product to be free from significant deviations in materi

Seite 34

TABLE OF CONTENTS 1 INTRODUCTION ...

Seite 35 - - 30

11 Appendices 11.1 Appendix 1 – Unit fuse and jumper settings Caution! Before opening the unit to replace the unit fuse, make sure the mains c

Seite 36 - - 31

11.2 Appendix 2 – pH buffer values at various temperatures The following table shows the various pH values at different temperature of the s

Seite 37 - - 32

11.3 Appendix 3 – Simple explanation on the function of hysteresis SP1 Set to LO SP2 Set to HI RELAY ONRELAY OFF4.59.5 10.04.0 The contro

Seite 38 - - 33

11.4 Appendix 4 – General instructions concerning Controller Setting 11.4.1 Control characteristic of Controllers used as limit value swit

Seite 39 - - 34

11.4.4 Control signal of Pulse length Controllers RelayONtONtOFFPulse Length TOFFTime [s] The output relay of the pulse length controller is cl

Seite 40 - 11 Appendices

11.4.5 Control signal of Pulse Frequency Controllers RelayONtONtOFFPulse Length TOFFTime [s] The output relay of the pulse frequency controlle

Seite 41

- 41 - 11.5 Appendix 5 – Abbreviations used in menu displays Abbreviation Meaning Abbreviation Meaning MEAS Measurement CUR. 1 O

Seite 43 - MIN function MAX function

Water Analysis InstrumentsNorth America166 Cummings CenterBeverly, MA 01915 USAToll Free: 1-800-225-1480Tel: 1-978-232-6000Dom. Fax: 1-978-232-6015Int

Seite 44

10 GENERAL INFORMATION...

Seite 45

INTRODUCTION 1.1 At the beginning We thank you for having purchased the Alpha pH 2000D pH/ORP Controller. The construction of the pH/ORP Control

Seite 46 - - 41

1.3 Safety Instructions - The Alpha pH 2000D should be installed and operated only by personnel familiar with the transmitter and who are quali

Seite 47

2 Product Description 2.1 Description of Unit The Alpha pH 2000D is used for measuring pH and temperature values. The pH values can be measured u

Seite 48 - Water Analysis Instruments

2.2 Measurement and Control System A typical measurement system consists of: • A pH/ORP process controller • A pH/ORP combination electrode w