Instruction Manual DO 110 Hand-held Dissolved Oxygen Meter with RS 232C TechnologyMadeEasy...68X361305 Rev. 0 - 11/05
Instruction Manual CyberScan DO 110 5NOTE: Do not substitute other probes or electrodes. For replacement probe, see the “Accessories” section.
Instruction Manual CyberScan DO 110 63.3 Connecting the AC/DC Adapter Besides using four AAA-sized batteries as a power source, the DO 110 met
Instruction Manual CyberScan DO 110 7 3.4.1 RS232C Configuration The DO 110 meter has a 9 pin female RS232C connector with the following pin o
Instruction Manual CyberScan DO 110 84 CALIBRATION 4.1 Important Information on Meter Calibration The amount of oxygen dissolved in water wi
Instruction Manual CyberScan DO 110 94.2 Temperature Calibration DO in mg/L is dependent on temperature, so it is first necessary to calibrate
Instruction Manual CyberScan DO 110 104.3 DO Calibration (with ATC) –% Saturation Mode Dissolved Oxygen exact calibration value depends on bar
Instruction Manual CyberScan DO 110 114.4 DO Calibration in mg/L (ppm) Mode The amount of oxygen dissolved in a liquid will depend on its temp
Instruction Manual CyberScan DO 110 125 MEASUREMENT During measurement, care must be taken not to allow the membrane of the DO probe touch any
Instruction Manual CyberScan DO 110 135.3 Taking Measurements 1. Rinse the probe with deionised or distilled water before use to remove any i
Instruction Manual CyberScan DO 110 146 MEMORY FUNCTION The DO 110 meter can store up to 100 sets of data. You can store data in any combinatio
Instruction Manual CyberScan DO 110 157 PRINT FUNCTION The DO 110 meter is equipped with an RS232 communication port and a PRINT () key that f
Instruction Manual CyberScan DO 110 167.2 Sending Data To Computer or Printer When the PRINT key is pressed, there are two options: • dAt /
Instruction Manual CyberScan DO 110 177.2.1 Print Current Data Manually To manually print any data that is currently being measured: 1. From
Instruction Manual CyberScan DO 110 187.3 Print Data from Stored Memory The DO 110 meter can print data that is stored in the memory. You can
Instruction Manual CyberScan DO 110 19To Print Selected Stored Data: 1. From measurement mode, press MR/▼ key. 2. “MEM” indicator appears a
Instruction Manual CyberScan DO 110 208 ADVANCED SETUP FUNCTIONS The advanced setup mode lets you customize your meter to suit your preference.
Instruction Manual CyberScan DO 110 21 8.1 Advanced SETUP Mode Overview Press the SETUP key to enter Setup mode. Press the MI/▲ and MR/▼ keys
Instruction Manual CyberScan DO 110 22 SETUP p 7.0 P7.0: Dissolved Oxygen Parameter P7.0 Select Barometric Pressure unit Set Ba
Instruction Manual CyberScan DO 110 238.2 P1.0 : Unit Configuration (COF) This subgroup program allows customising the meter to your specific
Instruction Manual CyberScan DO 110 248.2.2 P1.2: Select °C or °F temperature units The DO 110 meter lets you select between °C and °F units f
Preface This manual serves to explain the use of the hand-held series meters. The model covered is the DO 110 hand-held meter. The manual functions i
Instruction Manual CyberScan DO 110 25measurement mode. 8.2.4 P1.4: Select mg/L or ppm mode (Only in mg/L (ppm) Set Up mode) From the mg/L
Instruction Manual CyberScan DO 110 268.3 P2.0: Viewing Previous Calibration Data (CAL) This mode lets you recall previous calibration data,
Instruction Manual CyberScan DO 110 278.4.2 EL.2 Viewing the % Saturation Offset (Only available in % Saturation Set Up menu): This parameter
Instruction Manual CyberScan DO 110 288.5 P4.0 : Auto-Off (AtO) In Program P4.0 you can activate the AUTO-OFF option to automatically power of
Instruction Manual CyberScan DO 110 298.6 P5.0 : Reset to Factory Default (rSt) This mode lets you reset all parameters to factory default set
Instruction Manual CyberScan DO 110 308.6.2 User Reset This program clears all data including calibration, memory and other customised setup f
Instruction Manual CyberScan DO 110 318.7 P 6.0: Memory Clear (CLr) Use this parameter to clear all memory values when you need to store a ne
Instruction Manual CyberScan DO 110 328.8 P7.0: Dissolved Oxygen Parameters (DPr) This Set Up menu allows you to set the barometric pressure a
Instruction Manual CyberScan DO 110 338.9 P 8.0: % Saturation Offset Adjustment (OFS) This is a useful feature that allows you to offset meter
Instruction Manual CyberScan DO 110 349 CYBERCOMM PORTABLE DAS The DAS software is designed for pH 110, CON 110 and DO 110 meters to allow you
TABLE OF CONTENTS 1 INTRODUCTION ... 1 2 DISPLAY &a
Instruction Manual CyberScan DO 110 35 Figure 26: Click on 'Browse' button and locate CD-ROM drive Figure 27: Locate the CyberComm
Instruction Manual CyberScan DO 110 36 Figure 28: Select ‘Portable Meters’ directory to go to CyberComm Portable. Figure 29: Open ‘CyberComm
Instruction Manual CyberScan DO 110 37 Figure 30: Select "SETUP.EXE" program and click the OPEN button Figure 31: InstallShield Wiz
Instruction Manual CyberScan DO 110 38 Figure 33: Key in your name and company name and click NEXT button Figure 34: To select another Dest
Instruction Manual CyberScan DO 110 39 Figure 35: Creating a new program folder. Click on NEXT button. Figure 36: Click on NEXT button.
Instruction Manual CyberScan DO 110 40 Figure 37: The CyberComm Portable DAS program is fully installed. Click on FINISH button to end installa
Instruction Manual CyberScan DO 110 419.3 Running CyberComm Portable Before running the DAS program, please ensure that the RS232 cable is con
Instruction Manual CyberScan DO 110 42i. Buttons & Check-Box • Enable Connection - Click this button to enable communication between mete
Instruction Manual CyberScan DO 110 43 Figure 41: Communication Settings for computer's Com port. It must match with COM port settings on
Instruction Manual CyberScan DO 110 44 9.4 Capturing And Printing Data Into Computer Using Data Acquisition After matching the Communication
8.6.2 User Reset ... 30 8.7 P 6.0:
Instruction Manual CyberScan DO 110 4510 Dissolved Oxygen Probe 10.1 Dissolved Oxygen Principle The probe is a galvanic measuring element whi
Instruction Manual CyberScan DO 110 4610.3 Membrane Housing Replacement Replacement of the membrane cap housing/ membrane is required only whe
Instruction Manual CyberScan DO 110 47Figure 35: Use Tool to take out (or put in) membrane 10.4 Membrane/O-ring Replacement (Optional Procedur
Instruction Manual CyberScan DO 110 4814. Replace probe guard. 15. Calibrate the probe (see section 4) after the % saturation readings have
Instruction Manual CyberScan DO 110 4911 ERROR MESSAGES The following table provides a guideline to enable diagnosis of possible problems ind
Instruction Manual CyberScan DO 110 50then there is probably moisture somewhere. See Problem 2. With membrane cap removed and probe internals
Instruction Manual CyberScan DO 110 5113 LIST OF ACCESSORIES 13.1 Replacement Meter and Meter accessories Eutech Instruments Items Order Cod
Instruction Manual CyberScan DO 110 5214 SPECIFICATIONS Dissolved Oxygen Range Resolution Relative accuracy 0.00 – 19.99 mg/L or ppm 0.01 m
Instruction Manual CyberScan DO 110 5315 WARRANTY This meter is supplied with a three-year warranty, six-month warranty for probe against signi
Instruction Manual CyberScan DO 110 11 INTRODUCTION Thank you for selecting the DO 110 meter. These meters are microprocessor-based instruments
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website
Instruction Manual CyberScan DO 110 22 DISPLAY & KEYPAD FUNCTIONS 2.1 Display The LCD has a primary and secondary display. • The primary
Instruction Manual CyberScan DO 110 32.2 Keypad A large membrane keypad makes the instrument easy to use. Some buttons have several function
Instruction Manual CyberScan DO 110 43 PREPARATION 3.1 Inserting the Batteries This meter is packaged with 4 “AAA” alkaline batteries required